1. LEEP knife cutting: allows maximum power cutting, with a focus on cutting function. Using specialized LEEP scalpel heads to perform conization, circumcision, electrocoagulation, and hemostasis on the lesion site;
2. Radiofrequency ablation: also known as low-temperature plasma ablation, this technology is a new topic in the current international medical field, mainly used in the treatment of gynecological, uterine, and cervical diseases, with excellent results. The working temperature is relatively low (about 60-80 ℃), so it will not cause damage to normal tissues. The treatment mechanism is manifested as the resistance or insertion of the treatment electrode (or probe) into the lesion, using high-energy radio frequency waves to generate thermal effects on the polar molecules and ions in the lesion, using internal heating of the lesion, cell dehydration and atrophy, tissue protein degeneration and coagulation, sterile necrosis and shedding, achieving good therapeutic effects.
2. Performance indicators
1. Equipped with an automatic voltage stabilization system (ensuring stable terminal output when the external voltage is unstable);
2. LEEP cutting function: 5-100W, continuously adjustable power, with a step size of 5W;
3. RF ablation function: 0-60W, continuously adjustable power, with a step size of 5W;
4. Electrocoagulation: 0-60W, continuously adjustable power, with a step size of 5W;
5. Power display mode: LED digital display;
6. Power adjustment method: waterproof button adjustment;
7. Parameter settings: able to save commonly used cutting power and electrocoagulation power.
3. Handling performance
1. It has an automatic compensation function for output power, and the P.A.C system automatically adapts to various human body impedances;
2. Equipped with manual or foot controlled power output function;
3. Accurate positioning and superior hemostatic performance;
4. It has power memory function.
4. Safety performance
1. It has dual circuit safety automatic monitoring, control (automatic monitoring of abnormalities and shutdown of power output) and alarm functions, and can provide sound and light alarm prompts;
2. Automatic detection and elimination of high-frequency leakage current;
3. Unique use of internationally advanced N. E. F. Cooling method has changed the commonly used fan style cooling method by domestic electric knife manufacturers, avoiding secondary pollution in the operating room and advancing the development trend of the industry in the future.
5. Advantages
1. Digitalization and multifunctionality;
2. The surgical incision is smooth and fine, with no postoperative edema and no fibrous contraction scars on the incision;
3. Control method:
(1) Manual button.
(2) Dual control of foot pedal, anti slip design, remote control of foot switch.
4. Micro computer control, operating under high safety conditions.
5. Built in relevant fault self-test software, automatically performs self-test work after startup.
6. Main technical parameters of the machine
Main load frequency: 115KHz
Load impedance: LEEP cutting function 300 ?
RF ablation function: 100 ?
Working power supply: AC220V ± 10% 50Hz ± 2%
Maximum power consumption of the entire machine: ≤ 500W